        當前位置:首 頁 ? News ? Company News ? Blood pressure/pulse discrimination method

        Blood pressure/pulse discrimination method

        文章來源:鑫志科技 / 作者:鑫志科技 / 發表時間:2021-08-27

        Blood pressure/pulse discrimination method:

        The criterion of systolic pressure is certain: when the ratio of the amplitude Ui of a pulse wave to the maximum amplitude Um Ui/Um=Ks in the rising segment of the pulse wave amplitude envelope during inflation, the corresponding cuff pressure at this time is considered as the systolic pressure. Ps=P Å Ui=Ks · Um. The diastolic pressure criterion is determined: when the ratio of the amplitude Ui of a pulse wave to the maximum amplitude Um Ui/Um=Kd in the descending section of the pulse wave amplitude envelope, the corresponding cuff pressure at this time is considered as diastolic pressure. Namely, Pd=P Å Ui=Kd · Um. Heart rate refers to the number of beats per minute of the heart. Due to the divergence between the heartbeat and pulse, it is possible to measure heart rate while measuring blood pressure. The key to measuring heart rate is to distinguish the peak value of pulse waves, and then calculate the heart rate based on how many pulse waves there are within a certain period of time.

        Hardware circuit:

        Pneumatic parts:

        The pneumatic circuit consists of a cuff, an air pump, a pressure sensor, and a vent valve. The two signals output by the pressure sensor are filtered, amplified, and connected to the two A/D channels of the main controller to measure the static pressure signal and pulse signal; The air pump and vent valve are directly controlled by the main controller through the driving chip to complete the charging and discharging of the measurement time.

        Pressure sensing local:

        The characteristics of human physiological signals are low-frequency small signals, usually μ The V-mV level sensor has a frequency range of 0-300 Hz and a low signal-to-noise ratio. Therefore, it is very important to choose a sensor with good performance. In this plan, the US9111 sensor is selected to cooperate with two carefully designed filtering and amplification circuits to achieve high-quality signals.

        Filter and amplify local:

        The signal from the pressure sensor is a mixed signal of pulse signal and static pressure signal, mixed with high-frequency interference from the outside and DC or low-frequency weight. Static pressure signal belongs to low-frequency signal, with a frequency less than or equal to 0.04 Hz, and pulse signal frequency is generally about 1 Hz. Divide the mixed signal into two parts, one of which passes through a low-pass filtering circuit and is connected to an A/D converter to stop analog-to-digital conversion, obtaining static pressure signal data; The other path passes through another filtering and amplification circuit and stops analog-to-digital conversion to obtain amplified pulse signal data.

        咨詢

        QQ

        在線客服

        服務時間:9:00-18:00

        電話

        服務咨詢熱線:

        0755-2780-8770

        微信

        二維碼

        微信咨詢

        郵箱

        電子郵箱

        xuzb@xzkj.net.cn

        主站蜘蛛池模板: 精品国产三级a∨在线欧美| 午夜精品美女自拍福到在线| 国产一区二区精品久久凹凸| 精品国内自产拍在线观看| 精品无码国产污污污免费网站国产| 日韩精品一区二区三区中文字幕| 在线亚洲欧美中文精品| 18国产精品白浆在线观看免费| 午夜精品在线观看| 四虎影视永久在线精品| 久久久久久夜精品精品免费啦| 久久精品国产亚洲沈樵| 亚洲国产精品无码久久青草| 国内精品91最新在线观看| 国内精品久久久久久中文字幕| 亚洲国产精品成人精品无码区在线| 国产成人精品日本亚洲| 久久久久这里只有精品| 国产精品99久久精品| 亚洲愉拍99热成人精品热久久 | 精品无码人妻一区二区三区不卡| 无码人妻精品一区二区| 成人午夜精品久久久久久久小说| 亚洲Av永久无码精品三区在线| 国产精品青青在线观看爽香蕉| 久久精品国产第一区二区三区| 精品无码久久久久久久久久| 欧美精品在线一区二区三区| 久久精品www人人爽人人| 亚洲精品动漫免费二区| 黑人无码精品又粗又大又长| 欧美日韩精品一区二区三区| 国产亚洲欧美精品永久| 欧美成人精品高清在线观看| 亚洲码国产精品高潮在线| 国产色婷婷精品综合在线| 欧美激情精品久久久久久久九九九| 最新国产成人精品2024| 国产精品丝袜黑色高跟鞋| 一级A毛片免费观看久久精品| 十八18禁国产精品www|